CANDELA MINI GENTLEYAG MGY LASER HEAD, CAVITY, NEW , NO NEED CORE EXCHANGE.7121-00-7120 CALH1470

CANDELA MINI GENTLELASE MGL LASER HEAD, CAVITY, NEW , NO NEED CORE EXCHANGE

Candela GentleLASE, GentleYag Laser Head. 7121-00-7120

CALL FOR PRICING

1-866-444-8883

Description

Parts4Laser guarantees all of our products to be of the highest quality at very competitive prices.

Candela Mini GentleYAG MGY laser heads are packaged and ready for fast delivery. No need to send in your old laser head for a core exchange.

CALH1470 (7121-00-7120)

This is a new laser head, not a refurbished or rebuilt unit. All parts are brand new, including new metal parts and all new Flash Lamps, Triple Bore Flow Tube, Reflector, Light Blocking Hardware, and O-ring Seals.

These laser heads are all individually performance-tested in the factory. As a result, they show excellent output energy.

New laser heads greatly increase the useful life of the entire laser system, substantially lowering the voltages required for the same output energies.

Laser rod is NOT included.

We also offer installation of your laser rod and testing, laser rod refurbishing, and laser rod rework. For example, this service can help in the following situations:

  • When your system setting cannot achieve the setting and the power you request.
  • When you hear a thumping sound or when you do not get enough speed and energy during treatment.

This new laser head includes:

  • New metal and new water manifold
  • New flash lamps
  • New triple bore
  • New seal kit, including all o-rings
  • New light blocking kit
  • New flow tubes
  • Applying and re-coating with reflective powder

This Laser Head Fits the Following Machines

GentleLASE systems – GL / GL Plus
Mini GentleLASE
Mini GentleYAG
GentleMAX

We can also supply Yag and Alex laser heads and other laser rods if needed.

http://parts4laser.com/shop/candela-laser-gentlelase-pro-rod-755-alexandrite/

Candela GentleLase Laser Head Information

There are several reasons why the laser head can go bad. First, a water leak in the laser head can cause pressure to build up and break the laser rod. The rod is the device that channels the laser, so even one crack can make the laser head stop working.

Secondly, there is a reason why we are adamant about keeping the laser in a spotless-clean environment. If there is a small speck of dust on the optics, the laser will make contact with the dust and gradually burn the optic until the laser has nowhere to go except backwards or sideways. As a result, additional components can break. Additionally, dust can also burn the laser rod and even the laser head itself. It is also important to note that the flow tubes or triple bore can be damaged from water pressure or dust.

Lastly, a Candela GentleLase Pro laser head can malfunction or stop working altogether because of a damaged or broken flash lamp. Therefore, please pay attention to your lamp life. If you notice your power has decreased significantly, and you need more treatments to achieve the same result, it may be time to change your flash lamp.

In order to extend the life of your laser machine, we recommend a preventative maintenance service on your machine every six months. This helps ensure the safety of your patients, your investment, and your business, since a broken machine can cause delays and an inability to follow through on advertised services or scheduled appointments.

Parts4laser is not affiliated, associated, authorized, endorsed by, or in any way officially connected with Candela their affiliates. Any logos and trademarks of third parties that may be found on our website are registered trademarks of their respective owners. The use of any trademark is for identification and reference purposes only and does not imply any association with the trademark holder.

Parts4laser distributes directly to end users, medical spas, and dermatology and plastic surgery practices.

By purchasing this item, you agree to our terms and conditions.