CUTERA COOLGLIDE VANTAGE

The CoolGlide (1064 Nd:YAG) platform is the first laser cleared by the FDA for permanent hair reduction in all skin types. All skin types (I-VI) can be treated including tanned skin in the summer. This used cosmetic laser has Patented PowerFlex™ technology which provides fast treatments and high efficacy. Copper cooling is the most efficient method of cooling for optimal skin protection and comfort. Large spot size (10 mm) allows for maximum speed and optimal penetration at all follicle depths. The CoolGlide Platform supports three technologies and is an economical solution that will allow you o increase revenues while delivering multiple treatment programs to patients.

Handpiece Options: CoolGlide, Excel, Laser Genesis

Cutera Coolglide Applications

  • Hair removal on all skin types (including tanned skin)
  • Vascular treatments
  • Skin rejuvenation.
Category: Tags: ,

Description

Cutera Coolglide Vantage Aesthetic Laser Machine.

General Condition: Excellent

Manuals: Yes

Warranty: Yes

Type: 1064 Nd:YAG

Fluency: Up to 100 J/cm2

Repetition Rate: Single shot to 2 Hz

Fitzpatrick Skin Type Fluence Pulse Duration Spot Size
I-II 50-65 J/cm² 10-20 ms 10 mm
III 45-60 J/cm² 15-25 ms 10 mm
IV 40-55 J/cm² 15-30 ms 10 mm
V 35-50 J/cm² 25-35 ms 10 mm
VI 20-45 J/cm² 30-60 ms 10 mm

Fluence and Pulse Duration are determined by skin type, hair color, hair thickness and hair density.

1. Determine Skin Type
a. Use Fitzpatrick Skin typing chart provided to determine appropriate settings
b. Caution – when treating darker skin types, additional pre-cooling and slow repetition rate (Hz) are
required to protect the epidermis.
2. Determine Hair Color
a. Darker Hair Color = Lower Fluence, Longer Pulse Duration
b. Lighter Hair Color = Higher Fluence, Shorter Pulse Duration
3. Determine Hair Thickness
a. Coarse Hair = Lower Fluence, Longer Pulse Duration
b. Finer Hair = Higher Fluence, Shorter Pulse Duration
4. Determine Hair Density
a. Dense Hair = Lower Fluence, Longer Pulse Duration.
i. Dense hair requires longer pre-cooling and slower repetition rate
ii. Hair density should decrease with subsequent treatments
1. Parameters may need to be adjusted as density decreases, skin type permitting

All our machines are sold after a FULL service and testing and comes with a 90-day warranty.

Parts4laser  is not affiliated, associated, authorized, endorsed by, or in any way officially connected with Cutera their affiliates. Any logos and trademarks of third parties that may be found on our website are registered trademarks of their respective owners. The use of any trademark is for identification and reference purposes only and does not imply any association with the trademark holder.